IH Inlets for TEOM Monitor

New industrial hygiene (IH) size-selective inlets are available from Rupprecht & Patashnick for workplace monitoring and research applications using the TEOM Series 1400a Ambient Particulate Monitor. The indoor inlets operate at a flow rate of 2 l/min, and provide size-selective sampling of the thoracic, ISO Respirable and fine particle size ranges, with D50 cut points of 10, 4.5 and 2.5 microns, respectively.

 
Foam Cartridge in
Thoracic Inlet (10 micron).
Foam Cartridge in ISO
Respirable Inlet (4.5 micron).
Foam Cartridge in
Fine Inlet (2.5 micron).

The IH inlets connect to the air tube of the TEOM Sensor Unit. Disposable foam cartridges developed by the Health & Safety Laboratory in Sheffield, UK, under contract for Rupprecht & Patashnick provide for the thoracic, ISO Respirable and fine industrial hygiene particle size cut points. The foam cartridges are easy to exchange (see photo at right), and provide the flexibility to switch easily from one particle size range to another. Replacement foam cartridges are available in packages of 10, 25 and 100.

The performance of the industrial hygiene inlets are shown below for the design flow rate of 2 l/min:

 